Z-1690-2013Class I

MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine

Cerenovus, Inc. / initiated 2013-06-07 / Terminated / root cause: design

Codman & Shurtleff, Inc. began a recall of MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended... on . FDA classified it as Class I and gives the reason as "Miscalibrated Fill Level Sensor may affect dosing level". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1690-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cerenovus, Inc.
Product
MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended for the intrathecal delivery of Baclofen or Morphine.
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Miscalibrated Fill Level Sensor may affect dosing level
Root cause tag
design
FDA root cause
Device Design
Action
Codman issued a Medical Device Correction Notification to Medical Facilities dated June 7, 2013 via FedX to inform them of the potential problem and a Worksheet (instructions) to Identify Pumps with Miscalibrated FLS during each patient's next scheduled refill session or sooner if patient is symptomatic. If a pump is identified as having a miscalibrated FLS, notify Codman Neuro using the same worksheet. In addition a sample letter was provided to Physi cans to provide to patients on the system. Follow the recommendations for managing patients who are identified as having a pump with a miscalibrated FLS. For additional information,contact your Codman Neuro representative, or contact Codman Neuro Clinical Support at 1-800-660-2660. Please report any malfunction or adverse event related to this device to Codman Neuro at 1-866-491-0974, Option 2.
Quantity in commerce
2 units US and 735 OUS
Distribution
Worldwide Distribution-USA (nationwide) including the states of FL, NC, NJ, and TX, and the countries of Austria, Belgium, Spain, Great Britain, Germany, Italy, Ireland, Luxemburg, Russia, Saudi Arabia, Portugal, Greece, Egypt, Switzerland, and Turkey.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1690-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1690-2013Class ITerminatedVoluntary: Firm initiated
Z-1691-2013Class ITerminatedVoluntary: Firm initiated