Z-1562-2013Class II

ARTISTE with syngo RT Therapist version 4.3

Siemens Medical Solutions USA, Inc. / initiated 2013-05-29 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTISTE with syngo RT Therapist version 4.3 Product Usage: The intended use of the SIEMENS branded ARTISTE family of linear accelerator systems is to deliver... on . FDA classified it as Class II and gives the reason as "The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1562-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTISTE with syngo RT Therapist version 4.3 Product Usage: The intended use of the SIEMENS branded ARTISTE family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure MV or In-Line kView images using the syngo RT Therapist 4.3.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an Urgent: Medical Device Correction Customer Safety Advisory Notice letter to all affected customers. The letter identified the affected product, problem, what preventive measures can the user take and actions to be taken. Customers were advised to perform the preventive measures provided and inform all affected personnel immediately. The letter states that Siemens is currently working on a new version of the software. Customers are asked to include the Customer Safety Advisory Notice in the System Owner Manual chapter "Safety Advisory Letters". For question call 925-602-8157.
Quantity in commerce
8 units
Distribution
Worldwide distribution
Product codes and lots
All units of the version 4.3, Part No. 08162815

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1562-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1562-2013Class IITerminatedVoluntary: Firm initiated