Z-1834-2013Class II

AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers...

Siemens Healthcare Diagnostics Inc. / initiated 2013-06-03 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1834-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is for in vitro diagnostic use only.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Product Correction letter dated 5/29/13 was sent on the week of June 3, 2013 to all the customers who received the AU5800 Clinical Chemistry Analyzer with Automation Ready (AR) Connection (All Software versions). The letter informed the customers of the problem identified and the actions to be taken. Customers were instructed to complete and return the enclosed Response Form within 10 days and to share the information with their laboratory staff. Customers with questions are instructed to contact the 24 hours Call Center (Hotline) at (800) 854-3633 in the US and Canada.
Quantity in commerce
76 units
Distribution
Worldwide Distribution, including Nationwide (US) and the countries of China, Italy, Japan, Canada, Turkey, United Kingdom, Singapore, Israel, and the Netherlands.
Product codes and lots
Model Numbers: AU5811-01, AU5821-01, AU5831-01, AU5841-01, AU5811-02, AU5821-02, AU5831-02, AU5841-02, AU5811-03, AU5821-03, AU5831-03, AU5841-03, AU5811-04, AU5821-04, AU5831-04, AU5841-04, AU5811-06, AU5821-06, AU5831-06, AU5841-06  Part Numbers: A94906, A94911, A94916, A91921 A94907, A94912, A94917, A94922 A94908, A94913, A94918, A94923 A94909, A94914, A91949, A94924 A94910, A94915, A94920, A94925  Serial Numbers: All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1834-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1834-2013Class IITerminatedVoluntary: Firm initiated