AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers...
Beckman Coulter Inc. began a recall of AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1834-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is for in vitro diagnostic use only.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An Urgent Product Correction letter dated 5/29/13 was sent on the week of June 3, 2013 to all the customers who received the AU5800 Clinical Chemistry Analyzer with Automation Ready (AR) Connection (All Software versions). The letter informed the customers of the problem identified and the actions to be taken. Customers were instructed to complete and return the enclosed Response Form within 10 days and to share the information with their laboratory staff. Customers with questions are instructed to contact the 24 hours Call Center (Hotline) at (800) 854-3633 in the US and Canada.
- Quantity in commerce
- 76 units
- Distribution
- Worldwide Distribution, including Nationwide (US) and the countries of China, Italy, Japan, Canada, Turkey, United Kingdom, Singapore, Israel, and the Netherlands.
- Product codes and lots
- Model Numbers: AU5811-01, AU5821-01, AU5831-01, AU5841-01, AU5811-02, AU5821-02, AU5831-02, AU5841-02, AU5811-03, AU5821-03, AU5831-03, AU5841-03, AU5811-04, AU5821-04, AU5831-04, AU5841-04, AU5811-06, AU5821-06, AU5831-06, AU5841-06 Part Numbers: A94906, A94911, A94916, A91921 A94907, A94912, A94917, A94922 A94908, A94913, A94918, A94923 A94909, A94914, A91949, A94924 A94910, A94915, A94920, A94925 Serial Numbers: All
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1834-2013 | Class II | Terminated | Voluntary: Firm initiated |