Z-0126-2014Class II
Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images
Eastman Kodak Company / initiated 2013-06-06 / Terminated / root cause: design
Carestream Health, Inc. began a recall of Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images on . FDA classified it as Class II and gives the reason as "Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0126-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 6, 2013 Carestream Health Inc. initiated a recall for the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch. Carestream Health Inc. will replace the power plug in all potentially impacted DRX-Revolution Mobile X-Ray Systems. Your Carestream service representative will contact you to answer any questions and schedule a visit to replace the plug of your Mobile X-Ray System as soon as possible. If you have any questions or concerns, or if you experience any issue with your plug prior to our visit, please contact the Carestream Customer Care Center at 1-800-328-2910, 8AM through 8pm Eastern Standard Time.
- Quantity in commerce
- 180 units
- Distribution
- Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.
- Product codes and lots
- Serial Numbers: 111, 114, 119-123, 125-126, 134, 142-143, 151-155, 161-169, 173-178,180-183, 188-192, 196-209, 211, 221-232, 235-249, 254, 261-270, 275-276, 279-286, 305, 312, 316-326, 330-337, 339-349, 353, 355, 357-365, 372-373, 375-376, 379-385, 387-388, 394-401, 403-405, 409-410, 412-415, 417, 420-421, 423, 431-435, 441-446, 451-461, 464-470, and 482-490
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0126-2014 | Class II | Terminated | Voluntary: Firm initiated |