Z-1948-2013Class II

Colorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4)...

Ortho-Clinical Diagnostics, Inc. / initiated 2013-06-06 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of Colorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4) 6802413 Product Usage: The CM/RT incubator processes... on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1948-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Colorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4) 6802413 Product Usage: The CM/RT incubator processes Microslide reagents with the following VITROS systems: 1) VITROS 5,1 FS Chemistry System, 2) VITROS 5,1 FS Chemistry System Refurbished, 3) VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), VITROS 5600 Integrated System.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Ortho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 6, 2013 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that an OCD trained service personnel will contact them and schedule a replacement of the CM/RT wear pads. OCD requested customers to complete and return the enclosed Confirmation of Receipt form no later than June 17, 2013. Customers are asked to place a copy of the notification by each VITROS System in their facility. Foreign affiliates were notified by email on June 6, 2013 of the issue and instructed to notify their consignees of the issue and actions. If you have any questions or need additional information, please call Customer Technical Services at 1-800-421-3311.
Distribution
Worldwide Distribution - USA Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Puerto Rico and the countries of Mexico, Singapore, Venezuela, England, Germany, Italy & Spain.
Product codes and lots
Serial Numbers: 1) J34000102 through J34002323, 2) 34000133, 34000193, 4000212, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557,  34000596, 34000601, 34000631, 34000642, 34000805, 34000833, 34000851, 34000893, 34000947, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274, 34001297, 34001311, 34001317, 34001345, 34001378, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001589, 34001610, 34001612, 34001633, 34001668, 34001809,  34001840, 34001850, 34001897, 3) J46000108 through J46000312, 4) J56000110 through J56001779.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1948-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1948-2013Class IITerminatedVoluntary: Firm initiated