Z-1691-2013Class I
MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine
Cerenovus, Inc. / initiated 2013-06-07 / Terminated / root cause: design
Codman & Shurtleff, Inc. began a recall of MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine Product Code: 91-4201US (40ml pump) 91-4201 (40ml pump)... on . FDA classified it as Class I and gives the reason as "Miscalibrated Fill Level Sensor may affect dosing level". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1691-2013
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cerenovus, Inc.
- Product
- MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine Product Code: 91-4201US (40ml pump) 91-4201 (40ml pump) Intended for the intrathecal delivery of Baclofen or Morphine.
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Miscalibrated Fill Level Sensor may affect dosing level
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Codman issued a Medical Device Correction Notification to Medical Facilities dated June 7, 2013 via FedX to inform them of the potential problem and a Worksheet (instructions) to Identify Pumps with Miscalibrated FLS during each patient's next scheduled refill session or sooner if patient is symptomatic. If a pump is identified as having a miscalibrated FLS, notify Codman Neuro using the same worksheet. In addition a sample letter was provided to Physi cans to provide to patients on the system. Follow the recommendations for managing patients who are identified as having a pump with a miscalibrated FLS. For additional information,contact your Codman Neuro representative, or contact Codman Neuro Clinical Support at 1-800-660-2660. Please report any malfunction or adverse event related to this device to Codman Neuro at 1-866-491-0974, Option 2.
- Quantity in commerce
- 5 US and and 360 OUS
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of FL, NC, NJ, and TX, and the countries of Austria, Belgium, Spain, Great Britain, Germany, Italy, Ireland, Luxemburg, Russia, Saudi Arabia, Portugal, Greece, Egypt, Switzerland, and Turkey.
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1690-2013 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1691-2013 | Class I | Terminated | Voluntary: Firm initiated |