Siemens Mammomat Inspiration - full-field digital mammography
Siemens Medical Solutions USA, Inc began a recall of Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography on . FDA classified it as Class II and gives the reason as "Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1694-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography
- Product code
- MUE Full Field Digital, System, X-Ray, Mammographic
- Reason for recall
- Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.
- Quantity in commerce
- 3
- Distribution
- United States Nationwide Distribution in the states of ND, OH and SC.
- Product codes and lots
- Model number 10140000, serial numbers 3118, 3243 and 3309
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1694-2013 | Class II | Terminated | Voluntary: Firm initiated |