Z-1694-2013Class II

Siemens Mammomat Inspiration - full-field digital mammography

Siemens Medical Solutions USA, Inc. / initiated 2013-02-28 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography on . FDA classified it as Class II and gives the reason as "Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1694-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.
Quantity in commerce
3
Distribution
United States Nationwide Distribution in the states of ND, OH and SC.
Product codes and lots
Model number 10140000, serial numbers 3118, 3243 and 3309

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1694-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1694-2013Class IITerminatedVoluntary: Firm initiated