510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1601-2013Class II

Brand Name: Multirate Infusor Devices. Indicated for the intravenous administration of medications

Baxter Healthcare Corporation / initiated 2013-06-07 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Brand Name: Multirate Infusor Devices. Indicated for the intravenous administration of medications on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1601-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Brand Name: Multirate Infusor Devices. Indicated for the intravenous administration of medications.
Product code
MEB Pump, Infusion, Elastomeric
Reason for recall
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality. Complaint rates for ruptures have since decreased. Baxter wants to ensure that no affected product remains in the field.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent Product Recall notification was sent to all affected consignees on 06/13/2013. The letter states the affected products, problem description, hazard involved and actions to be taken by customer/user, with contact information. The letter was accompanied by a Customer Reply Form and a table of Affected Product List. The recall notification instructs customers to segregate affected product and contact Baxter for instructions for return of product.
Quantity in commerce
3,255,096 total
Distribution
Distributed Nationwide and in Puerto Rico.
Product codes and lots
Product Codes: 2C1154KP, 2C1155KP, 2C9960KP, 2C9961KP  Lot Numbers: 10F030 10F066 10H043 10H104 10K027 10K099 10M090 10G028 10J059 10M019 10M089 10F058 10G069 10G042 10J048 10M026  Expiration Date: 06/30/13

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1601-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1600-2013Class IITerminatedVoluntary: Firm initiated
Z-1601-2013Class IITerminatedVoluntary: Firm initiated
Z-1602-2013Class IITerminatedVoluntary: Firm initiated
Z-1603-2013Class IITerminatedVoluntary: Firm initiated
Z-1604-2013Class IITerminatedVoluntary: Firm initiated
Z-1605-2013Class IITerminatedVoluntary: Firm initiated
Z-1606-2013Class IITerminatedVoluntary: Firm initiated