Z-1848-2013Class II

Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system

Siemens Medical Solutions USA, Inc. / initiated 2013-05-07 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system on . FDA classified it as Class II and gives the reason as "Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1848-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.
Root cause tag
software
FDA root cause
Software design
Action
Siemens initiated their correction of this product in May, 2013 by sending an Safety Update Instruction XP067/12S to resolve the software error to all affected customers. All units have been corrected. Further questions please call (610)-219-6300.
Quantity in commerce
4
Distribution
US Distribution including the states of CA, IN, NJ and OK.
Product codes and lots
Model number 10094200, serial numbers 4024, 4077, 4043 and 4051