Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system
Siemens Medical Solutions USA, Inc began a recall of Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system on . FDA classified it as Class II and gives the reason as "Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1848-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens initiated their correction of this product in May, 2013 by sending an Safety Update Instruction XP067/12S to resolve the software error to all affected customers. All units have been corrected. Further questions please call (610)-219-6300.
- Quantity in commerce
- 4
- Distribution
- US Distribution including the states of CA, IN, NJ and OK.
- Product codes and lots
- Model number 10094200, serial numbers 4024, 4077, 4043 and 4051
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1848-2013 | Class II | Terminated | Voluntary: Firm initiated |