510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2057-2013Class II

Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes...

Synthes (USA) Products LLC / initiated 2013-05-31 / Terminated / root cause: design

Synthes (USA) Products LLC began a recall of Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of... on . FDA classified it as Class II and gives the reason as "The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2057-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Product code
HRS Plate, Fixation, Bone
Reason for recall
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent an Urgent Notice: Medical Device Labeling Correction letter dated May 31, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm VA-LCP Two Column Volar Distal Radius Plate when used with a 1.2Nm Torque Limiting Attachment. A screw that passes through the VA-LCP plate is not likely to result in a serious adverse event although should it occur, additional medical intervention during the index procedure could be needed. If you have any questions, please call (610) 719-5450, or contact your Synthes sales Consultant.
Quantity in commerce
119, 891
Distribution
Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI SC, SD, TN, TX, UT, VA, VT, WA WI and WV.
Product codes and lots
Part number J8301-F

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2057-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2057-2013Class IITerminatedVoluntary: Firm initiated