Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary...
Synthes USA HQ, Inc. began a recall of Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary cavity on . FDA classified it as Class II and gives the reason as "Due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an introduction of foreign material at the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1952-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary cavity.
- Product code
- HTO Reamer
- Reason for recall
- Due to the coiled design for this product, the product is difficult to clean and the potential for corrosion on the device exists. The potential for harm exists if there is an introduction of foreign material at the operative site originating from the presence of corrosion and/or the retained undefined material that could not be adequately cleaned.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Sythes notified direct accounts by letter on 6/11/13 and requested them to check their inventory and remove affected lots from stock. Synthes requested that affected lots be returned and to call Synthes at 1-800-479-6329 for a Return Authorization Number. Completion and return of the Verification section at the end of the letter was requested, including negative responses.
- Quantity in commerce
- 13,050
- Distribution
- Nationwide Distribution.
- Product codes and lots
- All lots with part number 359.106, 359.107, 359.108, 359.109, 359.110, 359.111, 359.112, 359.113, 359.114, 359.115
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1952-2013 | Class II | Terminated | Voluntary: Firm initiated |