Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1745-2013Class II

Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector...

Synthes GmbH / initiated 2013-05-31 / Terminated / root cause: design

Synthes USA HQ, Inc. began a recall of Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone... on . FDA classified it as Class II and gives the reason as "Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1745-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone lengthening, including conditions such as congenital mandibular deficiencies or post-traumatic defects. The Multi Vector Distractor can be utilized to perform bone transport procedures as an alternative to bone grafts and free flaps.
Product code
DZL Screw, Fixation, Intraosseous
Reason for recall
Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however there
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent a Urgent Notice: Medical Device Removal letter dated May 31, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however, there is a remote potential that additional intervention would be required due to a non-functioning clamp. Synthes requested the product to be removed from inventory and returned to Synthes by calling 1-800-479-6329 to obtain a Return Authorization Number. Direct Accounts were asked to complete and return the Verification Section at the end of the letter indicating the quantity of affected product was found. Negative responses were also requested. If you have any questions please call (610) 719-5450.
Quantity in commerce
121
Distribution
US Nationwide Distribution including the states of MA, TX, MN, SC, UT, VA, and NY.
Product codes and lots
Titanium Multi Vector Distractor Pin Holding Clamp, body, limited bone stock Part # 487.941 with lot numbers: 3769720, 3779569, 3788731, 3794434, 7806093, 7823463 and Titanium Multi Vector Distractor Pin Holding Clamp, ramus, limited bone stock Part # 487.942 with lot numbers: 3769706, 3779571, 3788741, 7806096, 7823464.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1745-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1745-2013Class IITerminatedVoluntary: Firm initiated