Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver...

Baxter Healthcare Corporation / initiated 2009-07-31 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7... on . FDA classified it as Class II and gives the reason as "There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1678-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or subcutaneously for post-operative pain management.
Product code
MEB Pump, Infusion, Elastomeric
Reason for recall
There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Baxter sent an "Important Product Information" notification dated July 31, 2009, with a follow-up "Important Product Information" notification mailed on March 5, 2010. The notifications informed customers of the potential issue, and included instructions to ensure that the winged Luer cap is tightly secured after filling and priming. Customers were also instructed to follow normal clinical practices to reduce contact with any leaking medication, and to store pumps with pre-dispensed medication in a sealed pouch. If customers noticed a leak they were instructed to place the pump in a plastic bag, quarantine the device and call Baxter's Corporate Product Surveillance at 1-800-437-5176 for further instructions. Customers were asked to return the attached reply form via FAX to 1-847-270-5457. If customers further distributed the affected product they were instructed to forward the notification to other facilities. Customers with questions were instructed to call 1-800-933-0303. For questions regarding this recall call 224-270-4667.
Quantity in commerce
Code 2C1700KP: 7,128 units; Code 2C1702KP: 105,732 units
Distribution
Worldwide Distribution - USA (nationwide), Puerto Rico and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Czech Republic, Denmark (including Faroe Islands), Ecuador, Finland, France (including Martinique and Reunion Islands), Germany, Greece, Iceland, India, Ireland, Isreal, Italy, Luxemburg, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweded, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
*** 1) Product Code: 2C1700KP; Lots: 09C013, 09D023;  *** 2) Product Code: 2C1702KP; Lots: 08M005, 08M056, 08N055, 09A029, 09B008, 09B061, 09C015, 09D013, 09D043, 09D081, 09E041, 09E070

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1678-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1677-2013Class IITerminatedVoluntary: Firm initiated
Z-1678-2013Class IITerminatedVoluntary: Firm initiated
Z-1679-2013Class IITerminatedVoluntary: Firm initiated