510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1820-2013Class I

LPS Lower Extremity Dovetail Intercalary component. Intended for use in replacement of the mid-shaft portion...

Depuy Orthopaedics Inc. / initiated 2013-07-11 / Terminated / root cause: design

DePuy Orthopaedics, Inc. began a recall of LPS Lower Extremity Dovetail Intercalary component. Intended for use in replacement of the mid-shaft portion of the femur on . FDA classified it as Class I and gives the reason as "The LPS Lower Extremity Dovetail Intercalary component has the potential for fracture of the female component when exposed to certain physiological loads". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1820-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
LPS Lower Extremity Dovetail Intercalary component. Intended for use in replacement of the mid-shaft portion of the femur.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The LPS Lower Extremity Dovetail Intercalary component has the potential for fracture of the female component when exposed to certain physiological loads.
Root cause tag
design
FDA root cause
Device Design
Action
On 7/11/13, URGENT MEDICAL DEVICE RECALL NOTIFICATIONS were sent to consignees: Depuy Distributors, hospitals, and surgeons. End-user Hospital Customers will submit reconciliation forms to DePuy. Possible clinical implications of device failure are stated in these notifications, including poor mechanics and loss of function, pain, component dislocation, adverse tissue reaction, bone damage, potential need for revision surgery and the risks associated with this type of surgery. A patient letter will be included with the surgeon letter to assist in surgeon's notification and discussion with patients. The devices will be returned to DePuy Warsaw through the normal DePuy Returns process, to attention of Returns and marking H13-13 on the outside of the box. If a patient presents with a fractured LPS" Lower Extremity Dovetail Intercalary component with well-fixed proximal and distal stems and the surgeon determines LPS" Lower Extremity Dovetail Intercalary component is the best treatment option, DePuy plans to make the LPS" Lower Extremity Dovetail Intercalary component available through authorized approvals. Directions on how to purchase the component are provided in the Recall Communications. Any questions or concerns about the recall, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8am-5pm EDT)
Quantity in commerce
58 U.S. / 39 OUS
Distribution
Worldwide Distribution-USA including the states of FL, GA, ID, KS, KY, MI, MO, NC, NV, NY, OH, OK, PA, TN, TX, VA, and WA and the countries of AUSTRALIA, CANADA, ENGLAND, GERMANY, INDIA, COUNTY CORK, IRELAND, and TURKEY.
Product codes and lots
CATALOG NO. : Part Number: 1987-19-055 and lot numbers:130896, 132131, 132133, 209461, 209466, 219843, 232972, 295965, 310189, 336843, 349283, 352878, 374125, 379089, A1TAR1, A2AGN1, A2AGS1, A2AGT1, A2AGV1, B2VAL1, B43G11, B43G1A, B43G1B, B43G1C, B43G1D, B69HV1, B69HVA, BK1BB1, BW7DX1, BW7DX1A, E3SJD1, and EN4KJ1.  In certain documentation, these alphanumeric lot numbers may appear with three zeros at the end (e.g. A1TAR1 may appear as A1TAR1000).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1820-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1820-2013Class ITerminatedVoluntary: Firm initiated