syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo Dynamics system is intended for...
Siemens Medical Solutions USA, Inc began a recall of syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo Dynamics system is intended for the acceptance, transfer, display, storage, archive... on . FDA classified it as Class II and gives the reason as "Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1859-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo Dynamics system is intended for the acceptance, transfer, display, storage, archive, and manipulation of digital medical images, including quantification and reportgeneration. syngo Dynamics is not intended to be used for reading mammography images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Customer Safety Advisory Notice dated June 4, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1) With sDIS implemented, use the DICOM Modality List (DMWL), without sDIS matching rules enabled OR 2) Do not implement sDIS. Update instructions were also provided. The reported issue will be resolved in a modification to syngo Dynamics version 9.5 and syngo Dynamics VA10A which will be released in the near future. We regret any inconvenience that this may cause , and we thank you in advance for your understanding. For further questions please call (610) 219-6300.
- Quantity in commerce
- 9
- Distribution
- US Distribution only including the states of PA, OH, UT, CA, NY, MO, and TX.
- Product codes and lots
- Model Number: 10091637, 10091805, with serial numbers 85164, 85289, 85296, 85297, 85149, 85116, 85177, 85051, and 85278.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1859-2013 | Class II | Terminated | Voluntary: Firm initiated |