Z-1795-2013Class II

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for...

Siemens Medical Solutions USA, Inc. / initiated 2013-06-05 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging on . FDA classified it as Class II and gives the reason as "Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1795-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
Root cause tag
design
FDA root cause
Device Design
Action
On 6/5/13, consignees were notified by letter of a "Field Safety Notice" and received instructions about how to avoid the potential risk of this issue while using the device. Siemens also rolled out a modification of AXIOM Aristos MX that will avoid these potential injuries in the future.
Quantity in commerce
401
Distribution
Nationwide Distribution.
Product codes and lots
Model Number 5895003 with serial numbers 1052 through 2500.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1795-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1795-2013Class IITerminatedVoluntary: Firm initiated