Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus
Siemens Medical Solutions USA, Inc began a recall of Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system on . FDA classified it as Class II and gives the reason as "During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1750-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens sent an Urgent Customer Information letter to their consignees. The letter identified the affected product, problem , update Instructions and actions to be taken. Siemens began to equip affected systems with a modified board assembly in January, 2013.
- Quantity in commerce
- 23
- Distribution
- United States Nationwide Distribution - including the states of: CA, CO, FL, IA, IL, IN, MA, MD, MI, NC, NY, PA, TX, VA and WA and Puerto Rico.
- Product codes and lots
- Material number 10094137 -- serial numbers 147515, 147500, 147528, 147514, 147516, 147526; material number 10094141 -- serial numbers 154408, 154406, 154453, 154423, 154430, 154425, 154429; material number 10280959 -- serial numbers 160813, 160823, 160812, 160800, 160431, 160814, 160807, 160817, 160816; material number 10094139 -- serial number 157907.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1750-2013 | Class II | Terminated | Voluntary: Firm initiated |