Z-1780-2013Class II

Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number...

Philips North America LLC / initiated 2013-07-08 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping... on . FDA classified it as Class II and gives the reason as "IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1780-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping 1.4 GHz Access Point provides wireless connectivity between IntelliVue devices (listed below) and the Philips IntelliVue Information Center. MX40 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity
Root cause tag
software
FDA root cause
Software Design Change
Action
Philips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377
Quantity in commerce
20105 units
Distribution
USA Nationwide Distribution
Product codes and lots
Serial numbers ranging from: US12700465 to US20823366

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1780-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1780-2013Class IITerminatedVoluntary: Firm initiated