Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number...
Philips Healthcare Inc. began a recall of Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping... on . FDA classified it as Class II and gives the reason as "IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1780-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping 1.4 GHz Access Point provides wireless connectivity between IntelliVue devices (listed below) and the Philips IntelliVue Information Center. MX40 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Philips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377
- Quantity in commerce
- 20105 units
- Distribution
- USA Nationwide Distribution
- Product codes and lots
- Serial numbers ranging from: US12700465 to US20823366
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1780-2013 | Class II | Terminated | Voluntary: Firm initiated |