Synthes Trauma Nail System. The devices are indication for bone fixation
Synthes USA HQ, Inc. began a recall of Synthes Trauma Nail System. The devices are indication for bone fixation on . FDA classified it as Class II and gives the reason as "Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2058-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Trauma Nail System. The devices are indication for bone fixation.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.
- Quantity in commerce
- 1,011,783
- Distribution
- Nationwide distribution.
- Product codes and lots
- Trauma Ex Nails, all lot numbers up to and including lot # 7072305 and Trochanteric Fixation Nails, all lot numbers up to and including lot # 7060897
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2058-2013 | Class II | Terminated | Voluntary: Firm initiated |