Z-2058-2013Class II

Synthes Trauma Nail System. The devices are indication for bone fixation

Synthes GmbH / initiated 2012-11-02 / Terminated / root cause: packaging

Synthes USA HQ, Inc. began a recall of Synthes Trauma Nail System. The devices are indication for bone fixation on . FDA classified it as Class II and gives the reason as "Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2058-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Trauma Nail System. The devices are indication for bone fixation.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.
Quantity in commerce
1,011,783
Distribution
Nationwide distribution.
Product codes and lots
Trauma Ex Nails, all lot numbers up to and including lot # 7072305 and Trochanteric Fixation Nails, all lot numbers up to and including lot # 7060897

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2058-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2058-2013Class IITerminatedVoluntary: Firm initiated