Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20...
Ecolab Inc began a recall of Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park... on . FDA classified it as Class II and gives the reason as "Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0948-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microtek Medical LLC
- Product
- Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park, Illinois
- Product code
- KKX Drape, Surgical, Antimicrobial
- Reason for recall
- Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The recall was conducted to the consumer/user level. The firm notified their direct consignees of the problem and the recall by written letter mailed on July 2, 2013. The notice requested the consignees to quarantine any recalled product on hand, and complete and return the response form by July 31, 2013. The consignees who further distributed the recalled product were asked to notify their consignees.
- Quantity in commerce
- 6,000 individual units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada.
- Product codes and lots
- Lot Numbers: D111571, D120421, D123471, D83111, DA90721, and DA90961
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0943-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0944-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0945-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0946-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0947-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0948-2014 | Class II | Terminated | Voluntary: Firm initiated |