Z-0948-2014Class II

Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20...

Microtek Medical LLC / initiated 2013-07-02 / Terminated / root cause: manufacturing process

Ecolab Inc began a recall of Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park... on . FDA classified it as Class II and gives the reason as "Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0948-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microtek Medical LLC
Product
Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park, Illinois
Product code
KKX Drape, Surgical, Antimicrobial
Reason for recall
Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The recall was conducted to the consumer/user level. The firm notified their direct consignees of the problem and the recall by written letter mailed on July 2, 2013. The notice requested the consignees to quarantine any recalled product on hand, and complete and return the response form by July 31, 2013. The consignees who further distributed the recalled product were asked to notify their consignees.
Quantity in commerce
6,000 individual units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada.
Product codes and lots
Lot Numbers: D111571, D120421, D123471, D83111, DA90721, and DA90961

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0948-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0943-2014Class IITerminatedVoluntary: Firm initiated
Z-0944-2014Class IITerminatedVoluntary: Firm initiated
Z-0945-2014Class IITerminatedVoluntary: Firm initiated
Z-0946-2014Class IITerminatedVoluntary: Firm initiated
Z-0947-2014Class IITerminatedVoluntary: Firm initiated
Z-0948-2014Class IITerminatedVoluntary: Firm initiated