Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1816-2013Class II

Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599...

Synthes GmbH / initiated 2012-10-12 / Open, Classified / root cause: packaging

Synthes USA HQ, Inc. began a recall of Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for... on . FDA classified it as Class II and gives the reason as "It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1816-2013
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes GmbH
Product
Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
Root cause tag
packaging
FDA root cause
Packaging
Action
Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329. If you have any further questions please call 610-719-5450.
Quantity in commerce
8
Distribution
US Distribution including the states of CA, MN, NC and TN.
Product codes and lots
Part Number 390.007, Lot Number 6800599

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1816-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1816-2013Class IIOngoingVoluntary: Firm initiated