Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599...
Synthes USA HQ, Inc. began a recall of Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for... on . FDA classified it as Class II and gives the reason as "It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1816-2013
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Synthes GmbH
- Product
- Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Synthes sent a Urgent Notice Medical Device Recall letter dated September 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is requesting that you please examine your inventory for products with the above Part and Lot number and remove them. Call Synthes at 1-800-479-6329. If you have any further questions please call 610-719-5450.
- Quantity in commerce
- 8
- Distribution
- US Distribution including the states of CA, MN, NC and TN.
- Product codes and lots
- Part Number 390.007, Lot Number 6800599