Z-2086-2013Class II

Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to...

Alphatec Spine, Inc. / initiated 2013-06-28 / Terminated / root cause: manufacturing process

Alphatec Spine, Inc. began a recall of Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical... on . FDA classified it as Class II and gives the reason as "The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2086-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alphatec Spine, Inc.
Product
Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
The firm, Alphatec Spine, sent an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" with response sheet dated June 28, 2013 via a combination of e-mail, Fedex, and telephone contact to its customers. The notification described the product, problem and actions to be taken. The customers were instructed to immediately return any and all affected lots of the illico Ti Cannulated Polyaxial Screw 7.5 mm x 40 mm implants that have been sent to them; if the device has previously been implanted, there is no further or additional risk post-implantation; immediately contact Alphatec Spine Customer Service Department - [email protected] on how to replace the implants; and complete and return the last page of the notification via mail to Alphatec Spine, Inc. 5818 El Camino Real, Carlsbad, CA 92008; email: sarrhenius@alphatecspine or fax to: (760) 431-0289. If you have any questions or need any additional information, please contact Director of Compliance at 760-494-9636 or email at [email protected].
Quantity in commerce
333 screws
Distribution
Worldwide distribution: US (nationwide) including states of:US states: AZ, CA, FL, GA, IL, MD, NJ, NY, OK, PA, TN, TX,and WA ; and countries of: Spain, France, Japan. International: Spain, France, Japan.
Product codes and lots
Part Number 73875-40. Lots Affected:  647289, 648923, 649244, 649556, 649813, 650040, 650262, 650589.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2086-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2086-2013Class IITerminatedVoluntary: Firm initiated