Z-1793-2013Class I

Optima NM/CT 640

GE Healthcare / initiated 2013-06-13 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Optima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer... on . FDA classified it as Class I and gives the reason as "GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1793-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Optima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
Root cause tag
design
FDA root cause
Device Design
Action
GE sent an Urgent Medical Device Correction letter dated June 17, 2013 to all affected customers. The letter identified the affected product, recommended that qualified service personnel maintain the equipment and that Preventative Maintenance procedures are executed according to labeling. Also the Safety Chapter Sections should be re-reviewed with personnel to ensure proper operation of the equipment. GE notified customers again on July 03, 2013, via Urgent Medical Device Recall letter (including confirmation of delivery for US customers) and follow-up telephone call. Customers were instructed to cease using the affected system until GE Healthcare Field Engineer can complete an inspection of the system and perform any necessary repairs at no cost. A GE Healthcare representative will contact the hospitals to arrange for the inspection.
Quantity in commerce
25
Distribution
Worldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Venezuela, Vietnam.
Product codes and lots
System ID: 0910263253, 0910263256, 00139NUC08, 00162NUC11, 030232NU15, 215728OPT640, 262574O640, 270956NM640, 410543NUC640, 715735OPT1, 717782640NM, 808485NM640, 815564O640, 816234D640, 817250NM640, 847872O640, 901516GO640, 915839NU01, 928214NMCT, A5366308, A5639613, BD003NU01, DK1054NM03, RTD0782, XM203401.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1793-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1787-2013Class ITerminatedVoluntary: Firm initiated
Z-1788-2013Class ITerminatedVoluntary: Firm initiated
Z-1789-2013Class ITerminatedVoluntary: Firm initiated
Z-1790-2013Class ITerminatedVoluntary: Firm initiated
Z-1791-2013Class ITerminatedVoluntary: Firm initiated
Z-1792-2013Class ITerminatedVoluntary: Firm initiated
Z-1793-2013Class ITerminatedVoluntary: Firm initiated