Z-1978-2013Class II

220/240V AC, Advanced Perfusion System 1

Terumo Cardiovascular Systems Corporation / initiated 2013-07-26 / Terminated / root cause: software

Terumo Cardiovascular Systems Corporation began a recall of 220/240V AC, Advanced Perfusion System 1 Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for... on . FDA classified it as Class II and gives the reason as "TCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1978-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
220/240V AC, Advanced Perfusion System 1 Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns TM or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
TCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that displays a "999" pressure reading and results in stopping a pump, this pressure display can be cleared and the pump re-started by disconnecting the pressure transducer from the system.
Root cause tag
software
FDA root cause
Software design
Action
Terumo sent an Urgent Medical Device Recall notification letter dated July 26, 2013 to all affected customers. The letter identified the affected product, reason for correction, potential hazards of the recall issue, the correction to be implemented, and customer instructions for replying to the recall notification. Terumo CVS is providing an addendum to be placed in the Operator's Manual at the beginning of Chapter 7, Modules. Questions or concerns are directed to Terumo CVS Customer Service at 1-800-521-2818, Recall Fax 1-877-275-6860. Customer Service hours are Monday to Friday 8am-6pm ET.
Quantity in commerce
1657
Distribution
Worldwide Distribution - US Nationwide in the states of :AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MSMT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and countries of AUSTRALIA, UNITED ARAB EMIRATES (UAE),Hong Kong,Indonesia, Singapore, Taiwan, Thailand,Philippines, CHILE,Malaysia, South Korea, BELGIUM,Japan,Mexico,CANADA
Product codes and lots
Catalog number: 801764

Enforcement reports