Eon Mini - 3721 (Charger)
St. Jude Medical began a recall of Eon Mini - 3721 (Charger). Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs... on . FDA classified it as Class II and gives the reason as "St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1739-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Eon Mini - 3721 (Charger). Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
- Product code
- LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
- Reason for recall
- St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- ST. Jude Medical sent an Important Medical Device Safety Information dated July 26, 2012 to affected customers. The letter identifed the affected product, problem and actions to be taken. For questions or concerns contact your ST. Jude Medical representative or St. Jude Medical Neuromodulation Division at 1-866-240-6741.
- Quantity in commerce
- 70,638 (combined total of the Eon and the Eon Mini systems)
- Distribution
- Worldwide Distribution - USA Nationwide Including Puerto Rico and countries of: Argentina, Australia, Belgium, Brazil, Canada, China, Chile, Colombia, Ireland, Israel, India, Malaysia, Mexico, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland and United Kingdom.
- Product codes and lots
- All serial numbers manufactured
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1738-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1739-2013 | Class II | Terminated | Voluntary: Firm initiated |