Z-1806-2013Class II

ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or...

Aesculap Inc / initiated 2013-06-18 / Terminated / root cause: manufacturing process

Aesculap, Inc. began a recall of ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical... on . FDA classified it as Class II and gives the reason as "Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1806-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical correction or stabilization of the spine. The Aesculap PEEK Spinal Implant System is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree lordosis. The device is labeled correctly, only the etching of the implant may be incorrect.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Aesculap notified customers via email and phone on June 18, 2013, and were instructed to check their inventory and return affected product to Aesculap. Aesculap believes that the risk to patient is low. For further questions please call 1-800-234-9179.
Quantity in commerce
9
Distribution
US Distribution including the states of NV and MI
Product codes and lots
Part no. SN038P, batch number 51915765

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1806-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1806-2013Class IITerminatedVoluntary: Firm initiated