Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system

Siemens Medical Solutions USA, Inc. / initiated 2013-03-04 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system on . FDA classified it as Class II and gives the reason as "Siemens issued a Customer Safety Notice to inform customers of a potential problem when using syngo.plaza. In some instances, it can happen that the laterality information in mammography images is not displayed and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1975-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Siemens issued a Customer Safety Notice to inform customers of a potential problem when using syngo.plaza. In some instances, it can happen that the laterality information in mammography images is not displayed and these images may not be properly aligned to mammography segments.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Siemens, notified their customers on March 4, 2013 and sent a "Customer Safety Advisory Notice" letter dated April 03, 2013. The letter described the product, problem and actions to be taken. The customers were instructed to always check whether the laterality information is displayed in the image text of mammography images; if not, the DICOM header is accessible and can be checked for the tag "Laterality" (0020, 0060). If a system is not able to provide the DICOM tag "Image Laterality" (0020, 0062), configuring the image text by adding the tag was proposed. If you have sold the device/equipment and it is no longer in your possession, forward this safety notice to the new owner of the device/equipment and inform us about the new owner of the device/equipment. This issue will be resolved in the syngo.plaza software version VA20D_HF04 which is planned to be released in 02/2013. Customers were also instructed to contact the local Siemens service office to plan the update. If you have any questions, contact the SYNGO CEO or Vice President Quality at +49 (9131) 84-4874 or email to: [email protected].
Quantity in commerce
6
Distribution
Nationwide distribution: FL, MA, KS, NY, OH, PA, TX and WI.
Product codes and lots
Model number 10592457, serial numbers 100219, 100201, 100473. Update dated 5/13/13 included the following additional serial numbers: 100177, 100504, 100174, 100181, 100473, and 100539.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1975-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1975-2013Class IITerminatedVoluntary: Firm initiated