Z-1945-2013Class II
Siemens AXIOM Artis zeego x-ray, angiographic system
Siemens Medical Solutions USA, Inc. / initiated 2013-05-13 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Siemens AXIOM Artis zeego x-ray, angiographic system on . FDA classified it as Class II and gives the reason as "Siemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1945-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens AXIOM Artis zeego x-ray, angiographic system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Siemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, Siemens, sent an "Important safety notice" by letter on May 13, 2013 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to immediately notify the service department to clarify what further measures need to be taken, if after carrying out the regular brake test, the system can only be moved at a reduced speed. If this device has been sold and is therefore no longer in your possession, send this safety notice to the new owner and provide the new owner to Siemens. Update instruction AX012/13/S was issued, and a software update will be released as AX006/13/S and AX007/13/S which will modify the system software so that it will not be possible for information to be overlaid. If you have any questions call +49 (9191) 18 0.
- Quantity in commerce
- 132
- Distribution
- Nationwide distribution: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, LA, MA, MD, MN, MO,MS, NC, NE, NH, NJ, NY, OH, PA, PR, SD, TN, TX, UT, VA, WA, and WI.
- Product codes and lots
- Model number 10280959
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1945-2013 | Class II | Terminated | Voluntary: Firm initiated |