Oxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mcg, 50 discs per cartridge, 5...

Remel Inc / initiated 2012-11-21 / Terminated / root cause: labeling and UDI

Remel Inc began a recall of Oxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mcg, 50 discs per cartridge, 5 cartridges per box. The firm name on the box label is... on . FDA classified it as Class III and gives the reason as "Boxes labeled as CIP5 may contain cartridges of AMP5". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0528-2014
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
Oxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mcg, 50 discs per cartridge, 5 cartridges per box. The firm name on the box label is Oxoid Ltd., Basingstoke, UK. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes.
Product code
JTN Susceptibility Test Discs, Antimicrobial
Reason for recall
Boxes labeled as CIP5 may contain cartridges of AMP5.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Thermo Fisher sent a Field Safety Notice dated November 13, 2012, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Customers were instructed to check stocks of the affected product if necessary and contact Customer Service or their local Oxoid supplier who will issue a replacement. The notice should be passed on to all who need to be aware within their organization. Customers were also instructed to discard any product labeled as AMP5. For questions, customers were instructed to call +44 (0)1256 841144.discard any product labeled AMP5. For questions regarding this recall call 913-895-4185.
Quantity in commerce
60/5/50-disc cartridge boxes
Distribution
Nationwide Distribution including AL, CA, CO, FL, IA, KS, LA, MA, MS, NE, NH, NJ, NY, OK, OR, RI, TN, TX, VA, WA WI, and Puerto Rico.
Product codes and lots
Lot number: 1220205, exp July 21, 2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0528-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0528-2014Class IIITerminatedVoluntary: Firm initiated