Z-1956-2013Class II

StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning...

Medtronic Navigation, Inc / initiated 2013-07-15 / Terminated / root cause: software

Medtronic Navigation, Inc. began a recall of StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgical trajectories in open and percutaneous procedures on . FDA classified it as Class II and gives the reason as "Medtronic Navigation, Inc. has initiated a field correction to their Framelink software version 5.4 which can be used on Medtronic Navigation StealthStation 57 and i7 systems or a Medtronic Planning Station due to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1956-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
StealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgical trajectories in open and percutaneous procedures.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Medtronic Navigation, Inc. has initiated a field correction to their Framelink software version 5.4 which can be used on Medtronic Navigation StealthStation 57 and i7 systems or a Medtronic Planning Station due to the potential of discrepancy between the exam image and the displayed measurement or overlay under certain conditions of use.
Root cause tag
software
FDA root cause
Software design
Action
A Field Safety Notice was sent on 7/15/13 to users informing them that the Framelink software version 5.4 has been revised to correct the software anomalies described in this letter. A new version will be installed on their system(s) by a Medtronic Navigation representative. Until the software upgrade has been installed on their system, do not use high resolution examinations (greater than 256 x 256). If users are not sure of the exam resolution being used, firm recommend contacting the hospital imaging department or local Medtronic Navigation representative for assistance. Medtronic representatives will visit all identified hospitals and upgrade all systems to version 5.4.1. This will account for all consignees.
Quantity in commerce
572
Distribution
Worldwide Distribution.
Product codes and lots
Lot number(s): Not applicable; All kits containing the Framelink Software version 5.4 are affected by this corrective action.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1956-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1956-2013Class IITerminatedVoluntary: Firm initiated