Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters...
Medtronic Inc. Cardiac Rhythm Disease Management began a recall of Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters into the left side of the heart on . FDA classified it as Class II and gives the reason as "Medtronic identified a specific labeling issue for specific lots of Mullins Transseptal Catheter Introducer Sheath, model 008551, in which packages labeled as Mullins 7F (French) may actually contain an 8F(French)...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1876-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular catheters into the left side of the heart.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Medtronic identified a specific labeling issue for specific lots of Mullins Transseptal Catheter Introducer Sheath, model 008551, in which packages labeled as Mullins 7F (French) may actually contain an 8F(French) dilator. Medtronic is requesting the return of potentially mis-labeled Mullins Transseptal Catheter Introducer 7 French Sheaths. Due to the differences in size, an 8F dilator will not
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Consignees were sent on 7/8/2013 a Medtronic, Inc " Urgent Medical Device Recall" letter dated 8 July 2013. The letter was addressed to Healthcare Professionals. The letter identified the problem and the product involved in the recall. Advised consignees to remove and quarantine the product from their inventory and return the product. Requested consignees to complete the attached Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612. If the product has been further distributed advised consignees to please notify them of the recall.
- Quantity in commerce
- 269
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and the states of AZ, AK, CA, DE, GA, IL, KY, MA, MI, MN, MS, NV, NY, NC, OH, PA, TN, and TX, and the countries of AUSTRALIA, BRAZIL, CHILE, INDIA, and SINGAPORE.
- Product codes and lots
- Lot Numbers: 50672012, 50676944, 50683123, 50693426, 50710622.