DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214...
DeRoyal Industries Inc began a recall of DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216... on . FDA classified it as Class II and gives the reason as "DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2113-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DeRoyal Industries, Inc.
- Product
- DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.
- Product code
- NKC Tubing/Tubing With Filter, Insufflation, Laparoscopic
- Reason for recall
- DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The firm notified customers via letters dated 7/15/2013 sent UPS and via DeRoyal Representatives. Customers were asked to notify DeRoyal of any remaining inventory to return for replacement product or credit. Non-responders will receive additional notifications. Effectiveness checks will be performed by contacting non-responders by phone.
- Quantity in commerce
- 506,632 sets in US; 17,854 sets International
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Dubai, Arab Emirates, Quebec, Canada, Japan, and Ireland.
- Product codes and lots
- All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2113-2013 | Class II | Terminated | Voluntary: Firm initiated |