Z-2083-2013Class II

BRAINLAB; FRAMELESS SRS QA TARGET POINTER Robotics is a device used to compensate rotational patient...

Brainlab, Inc. / initiated 2013-07-16 / Terminated / root cause: design

Brainlab AG began a recall of BRAINLAB; FRAMELESS SRS QA TARGET POINTER Robotics is a device used to compensate rotational patient misalignment (roll and pitch) in a linear accelerator... on . FDA classified it as Class II and gives the reason as "The Frameless SRS QA Target Pointer - Pointer Cap with the engraved cross hairs may become loose even when only a moderate force is applied to the Pointer Cap". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2083-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
BRAINLAB; FRAMELESS SRS QA TARGET POINTER Robotics is a device used to compensate rotational patient misalignment (roll and pitch) in a linear accelerator environment for stereotactic radiosurgery or radiotherapy procedures. The Frameless Radiosurgery Components are a device used for fixation, localization and repositioning of the patient's: head and neck; and head, neck, and shoulders, in a linear accelerator environment for stereotactic radiosurgery/radiotherapy procedures.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The Frameless SRS QA Target Pointer - Pointer Cap with the engraved cross hairs may become loose even when only a moderate force is applied to the Pointer Cap.
Root cause tag
design
FDA root cause
Device Design
Action
BrainLab sent an "FIELD SAFETY NOTICE / PRODUCT NOTIFICATION" dated July 9, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of the SRS QA Target Pointer. Consignees can instead use the Phantom Pointer feature until Brainlab can provide replacement hardware. Brainlab anticipates being able to provide all impacted facilities with replacement hardware by the end of July 2013. We sincerely apologize for any inconvenience and thank you in advance for your co-operation. If you require further clarification, please feel free to contact your local Brainlab Customer Support Representative. Customer Hotline: +49 89 99 15 68 44 or +1 800 597 5911 (for US customers) or by E-mail: [email protected]. Fax Brainlab AG: + 49 89 99 15 68 33
Quantity in commerce
6 units
Distribution
Worldwide Distribution - US Distribution only to AR., and the countries of Australia, Germany and Japan.
Product codes and lots
Part No. 43410

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2083-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2083-2013Class IITerminatedVoluntary: Firm initiated