510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1976-2013Class I

Stryker Oasys Midline Occiput Plate-1) Small 2) Medium 3) Large 4) Large Long 5) Mini. Allows for positioning...

Howmedica Osteonics Corp / initiated 2013-05-30 / Terminated

Stryker Spine began a recall of Stryker Oasys Midline Occiput Plate-1) Small 2) Medium 3) Large 4) Large Long 5) Mini. Allows for positioning directly on the midline keel of the occiput on . FDA classified it as Class I and gives the reason as "Stryker has received reports from customers indicating post-operative fracture of the pin that connects the tulip head to the plate body". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1976-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Oasys Midline Occiput Plate-1) Small 2) Medium 3) Large 4) Large Long 5) Mini. Allows for positioning directly on the midline keel of the occiput.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
Stryker has received reports from customers indicating post-operative fracture of the pin that connects the tulip head to the plate body.
FDA root cause
Other
Action
Stryker Spine sent a 'Urgent Medical Device Removal Letter'/'Customer Response Form' to Branch Managers, Agent Principals, Quality Contacts, Hospital OR Risk Managers, OR supervisor on 5/30/2013 via Fed Ex. Inventory should be examined and returned.The attached form should be completed and returned as well. Stryker Spine sent 'Urgent Medical Device Removal Letters' to Spinal Implant Surgeons on 6/20/2013 via Fed Ex.
Quantity in commerce
1536 units (US) 880 (Foreign)
Distribution
Nationwide Distribution.
Product codes and lots
K093670   Catalog numbers:  Small - 48551044 Medium - 48551045 Large - 48551046 Large Long - 48551047 Mini- 48551048  All lots of the above

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1976-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1976-2013Class ITerminatedVoluntary: Firm initiated