Z-1874-2013Class I

MRidium 1145 Dose Reduction System (DERS) drug library kit. (This is an option for the MRidium 3860+MR...

Iradimed Corporation / initiated 2013-07-01 / Terminated / root cause: software

Iradimed Corporation began a recall of MRidium 1145 Dose Reduction System (DERS) drug library kit. (This is an option for the MRidium 3860+MR Infusion Pump System). The MRidium 3860+MRI Infusion... on . FDA classified it as Class I and gives the reason as "The Dose Error Reduction System (DERS) can indicate an incorrect recommended value for the rate via a specific key sequence during the initial infusion setup. This can result in a risk of over-infusion or under-infusion...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1874-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Iradimed Corporation
Product
MRidium 1145 Dose Reduction System (DERS) drug library kit. (This is an option for the MRidium 3860+MR Infusion Pump System). The MRidium 3860+MRI Infusion pump are intended for general hospital or clinical use by medical professional whenever it is required to infuse patients with intra-venous or intra-arterial fluids before, during or after magnetic Resonance imagining (MRI) scans.
Product code
FRN Pump, Infusion
Reason for recall
The Dose Error Reduction System (DERS) can indicate an incorrect recommended value for the rate via a specific key sequence during the initial infusion setup. This can result in a risk of over-infusion or under-infusion if the infusion is started with this rate value.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
The firm, Iradimed Corporation, notified by phone and sent an "URGENT MEDICAL DEVICE RECALL NOTICE" via email dated July 1, 2013 all consignees/customers. The notice described the product, problem and action to be taken. The customers were instructed to identify the MRidium 3860+ MR Infusion Pumps which have the 1145 DERS Library card (AM05) installed. If these are not installed, no further action is required at this time; however, if they are installed-remove the 1145 DERS Library Card from the 3860 Pump using the instructions provided. Product will be reconditioned by installing a software upgrade that corrects this problem. The updated software (version 3.5.1) can be seen during the 3860+ Pump's power on cycle. Should your users have any questions following these instructions, please contact Iradimed Customer Service; 866-677-8022 (toll free) or 407-677-8022 ext. 100 or via e-mail - [email protected]
Quantity in commerce
176 individuals kits
Distribution
Worldwide distribution: US (nationwide) including states of: AL, CA, CO, FL, GA, IL, KY, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, RI, TN, TX, UT, VA, and WA; and countries of: Australia, Austria, Denmark, Epsom, France, Germany, London, Netherlands, Saudia Arabia, and United Kingdom.
Product codes and lots
Part number 1145 Lots: 4501,4510, 4538, 4587, 4596, 4675, 4690, 4705, 4738, 4748, 4960, 4970, 5001, 5065, 5104, 6164, 5221, 5240, 5349, 5361, 5517, 5737, 5764, 5881, 6006, 6151, 6170, 6180, 6252, 6470, 6583, 6806, 6881, 6984, and 7213.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1874-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1874-2013Class ITerminatedVoluntary: Firm initiated