Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005...
Stryker Craniomaxillofacial Division began a recall of Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015 (15 cc) HydroSet... on . FDA classified it as Class II and gives the reason as "Stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadvertent implantation of silica gel...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2054-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Craniomaxillofacial Division
- Product
- Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015 (15 cc) HydroSet Injectable HA Bone Substitute Sterile R/EO, nonpyrogenic, Product of Ireland Stryker Injectable Cement is a self-setting, calcium phosphate cement intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects as well as in the augmentation or restoration of bony contour in the craniofacial skeleton. The powder is mixed with the sodium phosphate/PVP solution provided in the kit and loaded into the empty cement delivery syringe to be applied directly to the site for void filling or augmentation.
- Product code
- GXP Methyl Methacrylate For Cranioplasty
- Reason for recall
- Stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadvertent implantation of silica gel granulate which could cause foreign body reaction including inflammation and might lead to the need for revision surgery.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Stryker sent an Urgent Medical Device Recall Notices with attached Product Correction Bulletin dated, July 30, 2013, describing the product, problem and possible adverse events associated with the recall were sent to all affected customers. Customers were instructed to pass the notice and bulletin to all individuals who may be using the device, provided instructions on procedures which would help mitigate risk and return the customer response form to our Customer Service Team by email, fax or letter. Email: [email protected] Fax: 1-877-648-7114 Address: 750 Trade Centre Way, Suite 200 Portage, MI 49002 Customers responses will be tracked and non-responders will be contacted. Further questions please call (877) 534-2464.
- Quantity in commerce
- 12,719
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries through Stryker Distribution: Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Ireland, Italy, Latvia, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Spain, Sweden, Switzerland, United Kingdom, Argentina, Australia, Brazil, Jordan, Kingdom of Saudi Arabia, Kuwait, Latin America (others), New Zealand, Russia, Singapore, South Africa, and United Arabian Emirates.
- Product codes and lots
- Part 79-43903, Lot: IC01116, IC01130, IC01204, IC01233, IC01238 Part 79-43905 Lot: IC01084, IC01085, IC01086, IC01093, IC01123, IC01150, IC01166, IC01167, IC01168, IC01180, IC01188, IC01189, IC01190, IC01198, IC01199, IC01200, IC01213, IC01214, IC01229, IC01230, IC01239, IC01240, IC01245, IC01246, IC01270 Part 79-43910 Lot: IC01087, IC01090, IC01091, IC01153, IC01154, IC01163, IC01164, IC01165, IC01179, IC01183, IC01184, IC01201, IC01202, IC01203, IC01231, IC01243, IC01244, IC01248, IC01249, IC01261, IC01264, Part 79-43015 Lot: IC01100, IC01114, IC01161, IC01162, IC01182, IC01211, IC01212, IC01227, IC01228, IC01241, IC01242, IC01252, IC01253, IC01254, IC01267 Part 397003 Lot: IC01092, IC01101, IC01102, IC01103, IC01104, IC01105, IC01106, IC01107, IC01115, IC01127, IC01131, IC01205, IC01220, IC01232, IC01234 Part 397005 Lot: IC01088, IC01094, IC01095, IC01096, IC01097, IC01098, IC01110, IC01111, IC01118, IC01119, IC01120, IC01132, IC01133, IC01134, IC01135, IC01136, IC01137, IC01147, IC01148, IC01149, IC01158, IC01159, IC01160, IC01171, IC01172, IC01181, IC01193, IC01194, IC01195, IC01196, IC01223, IC01224, IC01247Part 79-43910 Lot: IC01087, IC01090, IC01091, IC01153, IC01154 Part 397010 Lot: IC01099, IC01108, IC01109, IC01112, IC01121, IC01138, IC01139, IC01140, IC01141, IC01151, IC01152, IC01155, IC01156, IC01157, IC01173, IC01174, IC01191, IC01192, IC01197, IC01210, IC01216, IC01219, IC01225, IC01226, IC01250, IC01251, IC01263, IC01266, IC01301 Part 397015 Lot: IC01113, IC01122, IC01124, IC01125, IC01142, IC01143, IC01144, IC01145, IC01146, IC01169, IC01170, IC01177, IC01178, IC01185, IC01186, IC01187, IC01206, IC01207, IC01208, IC01209, IC01217, IC01218, IC01221, IC01222, IC01265
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2054-2013 | Class II | Terminated | Voluntary: Firm initiated |