Artis zee systems x-ray, angiographic system
Siemens Medical Solutions USA, Inc began a recall of Artis zee systems x-ray, angiographic system on . FDA classified it as Class II and gives the reason as "The firm became aware of a potential issue with Artis zee systems bearing model numbers 10094135, 10094137, and 10094141 equipped with a card collimator from a specific lot where a dislocation of the scatter radiation...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0172-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee systems x-ray, angiographic system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- The firm became aware of a potential issue with Artis zee systems bearing model numbers 10094135, 10094137, and 10094141 equipped with a card collimator from a specific lot where a dislocation of the scatter radiation filter inside the card collimator may occur.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens sent an letter to all affected customers in January 2013, notifying them of the potential issue and urging customers to make appointments with Siemens to have software update AX 027/12/S realized in order to resolve the problem. The software update was made available from January 22, 2013, on. Customers were also asked to pass along this information if the device was no longer in their possession. The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 17 unit(s) will consist of the following: 1. The firm sent out a customer notification letter in January 2013 notifying them of the issue and highlighting the approved procedures of the Corrective Action Plan. 2. Siemens directed in the customer notification letter, for the end-user to make an appointment with the Customer Service organization to have safety update AX 027/12/S realized. 3. As stated in the safety update AX 027/12/S and in the e-mail response the update to correct this issue will be provided to the customer at no expense. 4. Siemens indicated in the customer notification letter that the AX 027/12/S update will be made available from January 22, 2013 and on. 5. The AX 027/12/S update provides general information, work instructions and a completion certificate on how to correct the dislocated scatter radiation filter inside the collimator. You may proceed with implementation of the CAP (subject to the conditions noted above). For further questions please call (610) 219-6300.
- Quantity in commerce
- 17
- Distribution
- US Distribution including the states of AZ, KY, PA, IA WV, FL, MO, OK, CA, AR, MT, NY, KS, and CT.
- Product codes and lots
- model numbers 10094135, 10094137, and 10094141 equipped with a card collimator from a specific lot
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0172-2014 | Class II | Terminated | Voluntary: Firm initiated |