The Discovery NM/CT 570c CT Standalone Scan The Discovery NM/CT 570c hybrid system is a combination of a...

GE Healthcare / initiated 2013-04-10 / Terminated

GE Healthcare, LLC began a recall of The Discovery NM/CT 570c CT Standalone Scan The Discovery NM/CT 570c hybrid system is a combination of a cardiac dedicated nuclear imaging subsystem and... on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of an incident where the wrong part of the patient body was scanned during a CT standalone exam with the Discovery NM/CT 570c". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1967-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
The Discovery NM/CT 570c CT Standalone Scan The Discovery NM/CT 570c hybrid system is a combination of a cardiac dedicated nuclear imaging subsystem and LightSpeed GE CT subsystem. This system is intended for cardiac Nuclear Medicine (NM) diagnostic imaging applications, head and body CT diagnostic imaging applications, as well as attenuation correction and anatomical localization. The Discovery product is a SPECT-CT system. It consists of two back-to-back gantries, the NM Gantry, carrying solid state Cadmium Zinc Telluride (CZT) detectors, the CT Gantry, a single patient table, power distribution unit (PDU), NMV integrated power supply (IPS) , two operator consoles with two acquisition systems (Nuclear and CT), interconnecting cables and associated accessories (ECG gating, etc.).
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has recently become aware of an incident where the wrong part of the patient body was scanned during a CT standalone exam with the Discovery NM/CT 570c
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare will remedy the issue or bring the product into compliance with each applicable Federal standard. GE Healthcare has provided field modification instructions (FMI) to update software for the Discovery NM/CT570c hybrid system. The software update will correct the incorrect positioning of the table not prescribed by the operator. The FMI includes: A software upgrade with foreign language support to be installed in the affected models listed above, Labels (FMI label and packing label), Software upgrade workflow procedure with functional verification for all modes of operation, and Certification Report, Postponement of Medical Device Correction, Rejection of Medical Device Correction, or Refusal and/or Location Unknown Documentation. For further questions please call (262) 513-4122.
Quantity in commerce
10 Systems
Distribution
US Distribution in the state of CT.
Product codes and lots
Part numbers H3000SA UFC Cardiac SPECT System, H3001JA UFC Gantry, H3001JE UFC VCT Connecting Kit, H3001JF DISCOVERY NMV/CT 570C Connecting Kit.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1967-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1967-2013Class IITerminatedVoluntary: Firm initiated