Z-2045-2013Class II

Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular...

Medtronic, Inc. / initiated 2013-07-19 / Terminated / root cause: manufacturing process

Micro Therapeutics Inc, Dba Ev3 Neurovascular began a recall of Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with... on . FDA classified it as Class II and gives the reason as "Covidien Neurovascular has initiated the recall of several lots of Pipeline Embolization Device (PED) because the Instructions For Use (IFU) do not include the following contraindication: Patients in whom a pre-existing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2045-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-necked intracranial aneurysms in the internal carotid artery from the petrous to the superior hypophyseal segments.
Product code
OUT Intracranial Aneurysm Flow Diverter
Reason for recall
Covidien Neurovascular has initiated the recall of several lots of Pipeline Embolization Device (PED) because the Instructions For Use (IFU) do not include the following contraindication: Patients in whom a pre-existing stent is in place in the parent artery at the target aneurysm location. Additionally, the designation of aspirin and clopidogrel was missing from the following contraindication "
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Covidien sent an Customer Notification letter dated July 19, 2013, to all affected customers who received affected lots of Pipeline Embolization Device (PED). The customer notification letter informed the customers of the problems identified and the actions to be taken. Customer are instructed to reply and fax response form to Covidien Customer Service at (763) 398-7001. Customers are instructed that Customer Service will issue an RGA number for the return of the affected product back to Covidien. Customers are also instructed that upon receipt of a Returned Goods Authorization (RGA) number, to write the RGA number on the FEDEX slip and send the product back to: Covidien Attention: Product Recall 14005 13th Ave Plymouth, MN 55441 Prepaid and preaddressed FEDEX slips are enclosed for their convenience. Once Covidien receives the returned units, product replacement units will be shipped directly to customers. Local Covidien Neurovascular representative is also available to assist them in expediting the exchange of product. Customers with any questions, are instructed to call (949) 839-3700.
Quantity in commerce
62 units
Distribution
US Distribution including the states of AZ, KY, MO, NC, GA, WI, OR, FL, NY, CA, IL, TN, NV, TX, CO, MA, IA, UT and VA.
Product codes and lots
Model/Lot Numbers:  FA-71400-25 / 9747602  FA-77350-18 / 9749412  FA-77400-20 / 9759088 FA-71500-35 / 9747604  FA-77375-16 / 9744487  FA-77425-16 / 9752788 FA-77300-16 / 9757068  FA-77375-20 / 9757065  FA-77425-20 / 9744123 FA-77325-14 / 9759141  FA-77400-14 / 9757233  FA-77425-20 / 9752438 FA-77325-16 / 9753034  FA-77400-14 / 9768698  FA-77450-18 / 9753033 FA-77325-16 / 9757024  FA-77400-16 / 9743468  FA-77475-20 / 9752787 FA-77325-18 / 9759560  FA-77400-18 / 9751579  FA-77500-20 / 9750678 FA-77350-16 / 9752785  FA-77400-20 / 9755707

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2045-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2045-2013Class IITerminatedVoluntary: Firm initiated