Z-2095-2013Class II

PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT...

Siemens Medical Solutions USA, Inc. / initiated 2013-07-18 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083. Product Usage: The intended... on . FDA classified it as Class II and gives the reason as "Siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2095-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083. Product Usage: The intended use of the SIEMENS branded, ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might break off.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens released Updated Instruction containing the Medical Device Correction Field Safety Notice of Improvement letter, which was delivered by the Siemens Customer Service Group to all affected customers. The release of this UI TH002/12/S to distribute the update instruction began on July 18, 2013. This document includes completion protocol, which will be completed at each facility and returned to Siemens. A copy of the notification should be kept in the Digital Linear Accelerator System owners manual under Safety Advisory letters. UPDATE: a secnd update intruction, TH003/12/S has been released to begin the replcement of tabletops identified through the inspection carried out through TH002/12/S.
Quantity in commerce
179 active devices.
Distribution
Worldwide Distribution - USA Nationwide in the states of NJ, IN, PA, FL, OH, MN, MO, CA, AL, NE MI, LA. and countries of:Argentina, China, South Africa, Pakistan, Spain, Switzerland, Italy, Algeria, Croatia, Turkey, Canada, Germany, France, Poland, Japan, Russian Federation. Lebanon, Egypt, Slovakia, Latvia, Kenya, Brazil, Kuwait, Saudi Arabia, Greece, Belgium, Taiwan, Portugal.
Product codes and lots
ONCOR, PRIMUS and ARTISTE Lican systems with tabletop component: Tabletop, 550TxT, TT-A 08632296* Tabletop, 550TxT, TT-S 10046140

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2095-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2095-2013Class IITerminatedVoluntary: Firm initiated