Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0199-2014Class II

Cutting Bur Chart (part number 20-0019) utilized with the Electric and Pneumatic Motor Systems including...

The Anspach Effort, Inc. / initiated 2011-05-06 / Terminated / root cause: labeling and UDI

The Anspach Effort, Inc. began a recall of Cutting Bur Chart (part number 20-0019) utilized with the Electric and Pneumatic Motor Systems including: MicroMax Motor Systems, XMax Motor Systems, EMax2... on . FDA classified it as Class II and gives the reason as "Anspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0199-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
Cutting Bur Chart (part number 20-0019) utilized with the Electric and Pneumatic Motor Systems including: MicroMax Motor Systems, XMax Motor Systems, EMax2 Motor Systems, and Emax2Plus Motor Systems, all models and serial numbers. Product Usage: Indication for use is cutting and shaping bone
Product code
HSZ Instrument, Surgical, Orthopedic, Pneumatic Powered & Accessory/Attachment
Reason for recall
Anspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their regulations and compliance requirements that determined the claim that the device is capable of dissection of biometals, bioplastics and cement has been verified, but is is not part of th
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The firm, Anspach Effort, Inc., sent an "Urgent Medical Device Labeling Correction" letter dated June 27, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Destroy or delete any items in their inventory immediately and replace with the attached revised Operating Manual, Product Catalog and/or Cutting Bur Chart. 2) Complete and return the attached Customer Reply Form indicating disposition of the labeling materials and confirming their receipt of this letter, and fax to 1-800-327-6661. 3) If they distributed any of the products to other services or facilities, please forward this information as appropriate. For additional copies, contact Anspach Customer Support at (800) 327-6887.
Quantity in commerce
2,020
Distribution
Worldwide Distribution: USA Nationwide and the countries of: Egypt, India, Germany, Turkey, Canada, Saudi Arabia, Greece, United Kingdom, Cayman Islands, Kazakhstan, Republic of Korea, Taiwan, Hong Kong, China, Lebanon, Argentina, Denmark, Belgium, Netherlands, Mexico, France, Australia, Panama, Italy, Guam, United Arab Emirates, Israel, Polan, Venezuela, Austria, Hungary, Lithuania, Norway, Malaysia, Colombia, Costa Rica, Ecuador, Jamaica, Peru, Thailand, Italy, Czech Republic, Cameroon, South Africa, Portugal, Finland, Sweden, Nigeria, Bahamas, Cambodia, Dominican Republic, Switzerland, Chile, Singapore, Spain, Syrian Arab Republic, Jordan, Japan.
Product codes and lots
part number 20-0019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0199-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0198-2014Class IITerminatedVoluntary: Firm initiated
Z-0199-2014Class IITerminatedVoluntary: Firm initiated
Z-0200-2014Class IITerminatedVoluntary: Firm initiated