Z-1994-2013Class II

Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100...

Getinge USA, Inc. / initiated 2013-08-02 / Terminated / root cause: sterility

Getinge USA Inc began a recall of Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc... on . FDA classified it as Class II and gives the reason as "AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1994-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Getinge USA, Inc.
Product
Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121C, 132C, 134C and 135C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.
Product code
FRC Indicator, Biological Sterilization Process
Reason for recall
AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.
Root cause tag
sterility
FDA root cause
Other
Action
GETINGE sent an DEVICE CORRECTION NOTICE letter dated July 31, 2013, and response forms were sent to customers on August 2, 2013, via Certified Mail Return Receipt Requested. The letters instruct customers to immediately discontinue use, quarantine, and destroy the affected lot. In addition, they instruct customers to: Check the appropriate box and record the quantity of affected unused product on the enclosed Correction Response Form OR If if they do not have any lots of affected unused inventory check the box indicating such. Destroy the unused product in accordance with your institution's policy and instructions for use. Record the method of destruction on the Correction Response Form. Be sure to include the Facility Name, Address, Contact Name and Phone Number. Sign, date and e-mail the form back to [email protected] within 10 days or sooner of receiving this notice. Credit will be issued upon receipt of the Correction Response Form. Replacement product of unaffected lots is currently available. The warehouse shipping the affected lot is no longer being used (due to business decision, not as result of this recall). The current warehouse is maintaining the product in a controlled temperature segregated room away from any potential sterilants. If you have any questions regarding this notification please call 1-800-475-1400 ext. 5026
Quantity in commerce
Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)
Distribution
Worldwide Distribution - USA and the country of Canada.
Product codes and lots
LOT SR-430 (Expiration 02/06/2014)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1994-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1994-2013Class IITerminatedVoluntary: Firm initiated