510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1990-2013Class III

ConvaTec Flexi-Seal SIGNAL Fecal Management System, Rx Only

ConvaTec Limited / initiated 2013-06-25 / Terminated / root cause: packaging

Convatec Inc. began a recall of ConvaTec Flexi-Seal SIGNAL Fecal Management System, Rx Only Product Usage: For use to manage fecal incontinence through the collection of liquid to semi-liquid... on . FDA classified it as Class III and gives the reason as "ConvaTec notified their customers on 6/25/2013 regarding the manufacturing irregularity that occurred related to the packaging of certain lots of Flexi-Seal Signal FMS. In certain lots, the lot number convention was...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1990-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConvaTec Limited
Product
ConvaTec Flexi-Seal SIGNAL Fecal Management System, Rx Only Product Usage: For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medication.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
ConvaTec notified their customers on 6/25/2013 regarding the manufacturing irregularity that occurred related to the packaging of certain lots of Flexi-Seal Signal FMS. In certain lots, the lot number convention was duplicated on the packaging.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
ConvaTec issued a notification letters to their customers dated June 25, 2013 via UPS. The letter informed the customer of the issue with the related product and the action to be taken. If the customer has any question regarding the notification they can contact customer service at 1-800-422-8811.
Quantity in commerce
22,896 units
Distribution
US Nationwide Distribution
Product codes and lots
Flexi-Seal Signal FMS Kit - ICC Code: 418000 SAP Code #1262167 Lot# 12-FM-01 and Lot# 12-FM-02.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1990-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1990-2013Class IIITerminatedVoluntary: Firm initiated
Z-1991-2013Class IIITerminatedVoluntary: Firm initiated