Synthes 2.4 VA-LCP 2 CLMN VLR DSTL Radius PL 7H HD/3H Shaft/Right
Synthes USA HQ, Inc. began a recall of Synthes 2.4 VA-LCP 2 CLMN VLR DSTL Radius PL 7H HD/3H Shaft/Right Product Usage: Intended for fixation of complex intra-and extra-articular fractures and... on . FDA classified it as Class II and gives the reason as "The possibility exists that 2.4 mm VA-LCP 2 column distal radius plates from a specific lot are etched with "R" for right plate when they should be etched with "L" for left plate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2087-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes 2.4 VA-LCP 2 CLMN VLR DSTL Radius PL 7H HD/3H Shaft/Right Product Usage: Intended for fixation of complex intra-and extra-articular fractures and osteotomies of the distal radius and other small bones.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The possibility exists that 2.4 mm VA-LCP 2 column distal radius plates from a specific lot are etched with "R" for right plate when they should be etched with "L" for left plate.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Synthes GMBH sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their inventory and immediately remove the affected product from stock. Customers were advised to complete the Verification Section at the end of the letter. For questions call 610-719-5450 or contact your Synthes Sales Consultant.
- Quantity in commerce
- 11
- Distribution
- USA Nationwide Distribution in the states of FL, GA, NC, OH, KY, NJ, ID, IN, and MN.
- Product codes and lots
- Part No. 02.111.730, with Lot No. 3614792
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2087-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY