Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S...
Stryker Howmedica Osteonics Corp. began a recall of Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S 4.0x42mm TL 14mm, sterile Product Usage: The... on . FDA classified it as Class II and gives the reason as "Stryker received a report that the length of the cannulated screw asnis III, article # 604640 Lot R11482, was incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2099-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S 4.0x42mm TL 14mm, sterile Product Usage: The Asnis III Cannulated Screw System is intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Stryker received a report that the length of the cannulated screw asnis III, article # 604640 Lot R11482, was incorrect.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker sent an Urgent Medical Device Removal letter dated July 8, 2013 via Fed Ex to all affected customers. The letter identified the affected product, product issues, potential hazards, risk mitigation and actions to be taken. Customers were instructed to return all affected products available to Stryer Osteosynthesis or contact Stryer customer service and complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-251-3635. Questions can be directed to Stryker at 201-972-2100.
- Quantity in commerce
- 23 units
- Distribution
- USA Nationwide Distribution in the states of: FL. OH, TX, & VA
- Product codes and lots
- Catalog # 604640, Lot Code R11428 21 units Catalog # 604642S, Lot Code R09071 1 unit
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2099-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY