510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2099-2013Class II

Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S...

Howmedica Osteonics Corp / initiated 2013-07-08 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S 4.0x42mm TL 14mm, sterile Product Usage: The... on . FDA classified it as Class II and gives the reason as "Stryker received a report that the length of the cannulated screw asnis III, article # 604640 Lot R11482, was incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2099-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642S 4.0x42mm TL 14mm, sterile Product Usage: The Asnis III Cannulated Screw System is intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.
Product code
HWC Screw, Fixation, Bone
Reason for recall
Stryker received a report that the length of the cannulated screw asnis III, article # 604640 Lot R11482, was incorrect.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent an Urgent Medical Device Removal letter dated July 8, 2013 via Fed Ex to all affected customers. The letter identified the affected product, product issues, potential hazards, risk mitigation and actions to be taken. Customers were instructed to return all affected products available to Stryer Osteosynthesis or contact Stryer customer service and complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-251-3635. Questions can be directed to Stryker at 201-972-2100.
Quantity in commerce
23 units
Distribution
USA Nationwide Distribution in the states of: FL. OH, TX, & VA
Product codes and lots
Catalog # 604640, Lot Code R11428 21 units  Catalog # 604642S, Lot Code R09071 1 unit

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2099-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2099-2013Class IITerminatedVoluntary: Firm initiated