ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vitro diagnostic use in the...
Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vitro diagnostic use in the quantitative determination of folate in serum or red blood... on . FDA classified it as Class II and gives the reason as "Calibration failures and/or significant negative shifts in quality control (QC) and patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2171-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA Centaur Folate (500 tests), REF SMN 10331250, Cat No. 00203473 In vitro diagnostic use in the quantitative determination of folate in serum or red blood cells.
- Product code
- CGN Acid, Folic, Radioimmunoassay
- Reason for recall
- Calibration failures and/or significant negative shifts in quality control (QC) and patient results.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Siemens sent an Urgent Medical Device Correction letter on July 26, 2013, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. an Urgent Field Safety Notice was sent to all domestic customers via Federal Express. On July 25, 2013 this notice was sent to all OUS customers via hard copy, email, and/or fax as determined by each country's local regulations and procedures. These notices inform the customer that a portion of Folate ReadyPacks in kit lots ending in 218 are exhibiting calibration failures and/or significant negative shifts in quality control (QC) and patient results. If customers have an alternate lot of the ADVIA Centaur or ADVIA Centaur CP Folate assay, they are instructed that this lot should be used for future testing. However if no alternative lot is available, quality control (QC) results should be obtained for each individual ReadyPack as a QC or calibration is an indication of an affected pack. An affected pack must be replaced with a different Folate ReadyPack where a valid calibration and acceptable QC values are obtained. These notices should be reviewed by the facility's laboratory or medical director, though no look back is required. Customers were instructed that Fedex Return Receipts will be used to confirm that the customers received the UMDC. Customers with questions were instructed to contact their local Technical Support Provider or Distributor. For questions regarding this recall call 508-668-5000.
- Quantity in commerce
- 488
- Distribution
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada and Mexico.
- Product codes and lots
- 22298218
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2169-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2170-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2171-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2172-2013 | Class II | Terminated | Voluntary: Firm initiated |