Z-2059-2013Class II

Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics

Philips North America LLC / initiated 2013-08-08 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics on . FDA classified it as Class II and gives the reason as "The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2059-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.
Product code
KPR System, X-Ray, Stationary
Reason for recall
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare sent an Urgent Field Safety Notice letter dated July 24, 2013, to all affected customer. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed : When reading an image, a physician should not use the mirror icon for image side interpretation. 2. Clinical staff should follow the Instructions for Use and use lead letters to indicate body orientation of patient or equivalent electronic markers to provide Left/Right indications, during all X-ray image acquisitions. Philips will upgrade the system, which involves software and hardware upgrades of the system. The software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Philips will also provide updated Instructions for Use for the work spot and the system (including an addendum). If you need any further information or support concerning this, please contact your local Philips representative Technical Support Line at 1-800-722-9377. We apologize for any inconvenience this may cause and trust that this information is adequately addressing any concerns you may have.
Quantity in commerce
89
Distribution
Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, GA, MO, MS, NC, NH, NY, PA and TX., and the countries of Canada, Australia, China, Czech Republic, France, Germany, Greece, Italy, Netherlands, Portugal, Russia, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates and United Kingdom.
Product codes and lots
437765/SN09000009, 452854/SN10000003, 463879/SN10000016, 500643/SN12000006,504710/SN12000011, 511454/SN12000018, 519869/SN13000005, 524432/SN13000007, 500643/SN12000006, 436570/SN09000008, 511492/SN12000019, SN09020039, SN09010077,  438946/SN09000011, SN10002064, SN11020126, SN12020282, 516048/SN12000021, SN13020163, 524411/SN13000011, 409205/SN08000002, 438474/SN09000040, 448425/SN10000001, 449743/SN10000010, 461559/SN10000014, 493489/SN11000008, 496231/SN12000003, 434381/SN09000005, 456026/SN10000011, 465845/SN11000005, 471608/SN11000001, 481912/SN11000004, 508150/SN12000015, 524367/SN13000009, 524510/SN13000010, SN07020246, 485916/SN11000007, 428148/SN09000006, 423585/SN09000003, 500351/SN12000009, 506215/SN12000012, 520201/SN13000003, 409442/SN08000003, 419279/SN09000001, 518054/SN13000001, 460532/SN10000013, SN09020304, 443304/SN09000012, 443310/SN09000013, 416734/SN08000005, 420531/SN09000002, 423195/SN09000004, 460544/SN10000017, 454997/SN10000009, 496323/SN12000002, 496324/SN12000001, 399965/SN08000001, 438232/SN09000010, SN09020302, 443840/SN09000014, 452406/SN10000004, 454310/SN10000005, 478444/SN11000003, 502350/SN12000007, 502355/SN12000016, 520759/SN13000006, 434429/SN09000007, 451666/SN10000002, 453781/SN10000008, 453960/SN10000007, 459266/SN10000012, 460755/SN10000015,  466174/SN10000019, 468076/SN10000020, 471702/SN11000002, 499486/SN12000004, 501950/SN12000008, 503132/SN12000010, 507693/SN12000013, 507694/SN12000014, 519994/SN13000004, 520759/SN13000006,SN09001033, SN09020312, SN09000826, SN11000761, 493873/SN11000009, SN12020117, SN12020322, SN12020440

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2059-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2059-2013Class IITerminatedVoluntary: Firm initiated