UNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures
Aesculap, Inc. began a recall of UNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures on . FDA classified it as Class II and gives the reason as "Changes were implemented to the UNITRAC instructions for use to ensure safe use and potentially extend the life of the product. This included recommending the use of Sterile Cover JG901 and new Quick Release Coupling...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0031-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- UNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures.
- Product code
- GAD Retractor
- Reason for recall
- Changes were implemented to the UNITRAC instructions for use to ensure safe use and potentially extend the life of the product. This included recommending the use of Sterile Cover JG901 and new Quick Release Coupling RT02OR to eliminate the need to sterilize the UNITRAC and a functional check was added prior to use to ensure optimal functionality of the UNITRAC system.
- Root cause tag
- labeling and UDI
- FDA root cause
- Process change control
- Action
- A letter dated July 16, 2013 was sent requesting users to review the new instructions for use in their entirety, sign the attached Hospital Acknowledgment form and return to Aesculap. An Aesculap representative would be in contact to provide an in-service regarding the process for set-up of the Sterile Cover and required Functional Check. Discontinuation of use and repair of the device was recommended if the joints did not move freely or if the set position drifts during the Functional Check.
- Quantity in commerce
- 217
- Distribution
- Nationwide Distribution.
- Product codes and lots
- RT040R, all lot numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0031-2014 | Class II | Terminated | Voluntary: Firm initiated |