Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and...
Synthes USA HQ, Inc. began a recall of Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle on . FDA classified it as Class II and gives the reason as "A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0014-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by emphasizing the need for supplemental fixation.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- An "Urgent Notice: Medical Device Recall" letter dated March 8, 2013 was sent to customers asking them to check their inventory, remove all affected product technical guides from use, and return them to Synthes along with the Verification form. Customer questions were directed to (610) 719-5450.
- Quantity in commerce
- 62
- Distribution
- Nationwide Distribution, including the states of MI, PA, SC, IA, MA, CA, WI, IL, and KS.
- Product codes and lots
- Part Nos.: 04.111.210, 04.111.220, 04.111.230. with Lot Nos.; 2294209 through 7587547.