Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0014-2014Class II

Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and...

Synthes GmbH / initiated 2013-03-08 / Terminated / root cause: labeling and UDI

Synthes USA HQ, Inc. began a recall of Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle on . FDA classified it as Class II and gives the reason as "A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0014-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.
Product code
HWC Screw, Fixation, Bone
Reason for recall
A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by emphasizing the need for supplemental fixation.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
An "Urgent Notice: Medical Device Recall" letter dated March 8, 2013 was sent to customers asking them to check their inventory, remove all affected product technical guides from use, and return them to Synthes along with the Verification form. Customer questions were directed to (610) 719-5450.
Quantity in commerce
62
Distribution
Nationwide Distribution, including the states of MI, PA, SC, IA, MA, CA, WI, IL, and KS.
Product codes and lots
Part Nos.: 04.111.210, 04.111.220, 04.111.230. with Lot Nos.; 2294209 through 7587547.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0014-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0014-2014Class IITerminatedVoluntary: Firm initiated