Z-0024-2014Class II

GE PET Discovery 610, PET Discovery 710. Intended for head and whole body attenuation corrected Positron...

GE Healthcare / initiated 2013-03-26 / Terminated / root cause: software

GE Healthcare, LLC began a recall of GE PET Discovery 610, PET Discovery 710. Intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0024-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE PET Discovery 610, PET Discovery 710. Intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 26, 2013. The letter was addressed to Director of Radiology, Emergency Medicine Physicians and Radiologic Technologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information.
Quantity in commerce
28 (4 US & 24OUS)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, IS, LA, MA, MO, NH, NM, NY, OH, OK PA, SC, TX, UT, VA, WA, and WI, and the countries of UNITED KINGDOM, DENMARK, CHILE, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, CANADA, CZECH REPUBLIC, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, FRANCE GUIANA, GERMANY, GUADELOUPE, GUATEMALA, GUYANA, INDIA, INDONESIA, IRAQ, ITALY, JAPAN, JORDAN, KAZAHSTAN, KOREA, KUWAIT, LATVIA, MYANMA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, POLAND, ROMANIA, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, UZBEKISTAN, and VIETNAM.
Product codes and lots
MfgLotorSerial# 00000428413CN9 00000429077CN1 00000429247CN0 00000430152CN9 00000430656CN9 00000430999CN3 00000431369CN8 00000429701CN6 00000430064CN6 00000431187CN4 00000429354CN4 00000431447CN2 00000293950HM2 00000429821CN2 00000429946CN7 00000430001CN8 00000293950HM2 00000430632CN0 Shipped Shipped Shipped 00000429572CN1 00000428755CN3 00000428755CN3 00000428868CN4 00000293949HM4 00000428536CN7 00000429820CN4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0024-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0023-2014Class IITerminatedVoluntary: Firm initiated
Z-0024-2014Class IITerminatedVoluntary: Firm initiated